Stage travel, working distance and defect contrast decide whether a wafer microscope is usable at 6 inches or only at 12 — the optics are the easy part.
A 6-inch wafer needs at least 158 mm of X-Y travel to reach the edge without re-chucking; 8-inch needs 210 mm; a 12-inch wafer, photomask or panel needs 356 mm. Our three platforms are built around exactly those three numbers: STMT60, STMT80 and STMT12.
Wafers arrive on carriers, frames or chucks. Objectives with 15 mm working distance at 5× and 8+ mm at 10× clear the fixture and still deliver the NA needed for defect classification. Standard bright-field metallurgical objectives run out of clearance immediately.
Particles, residues, scratches and photoresist defects scatter light. BD semi-apochromat objectives with a coded turret let an inspector switch contrast without losing position — the defect stays in the field while the contrast changes.
A 12 MP industrial camera with a 0.65× adapter gives a field of view that matches what the eyepiece shows, so a defect photo is evidence rather than an approximation. Stitching and extended depth of focus in ST-1.5D cover features larger than one field.

A 710 × 420 mm stage with 356 × 305 mm travel — whole-wafer, photomask and panel review on one platform.…
View specifications
A 450 × 240 mm table with 158 × 158 mm travel — the 6-inch workhorse for compound-semiconductor and MEMS fabs.…
View specifications
525 × 330 mm table with 210 × 210 mm travel — 8-inch capacity without giving up ergonomics.…
View specificationsMaterial, size, feature of interest and the measurement you need to report.