Polarizing & mineralogical microscopy

Birefringence is a measurement of angle and retardation. The stage, not the objective, is the precision component.

Home / Applications / Polarizing & mineralogical microscopy
What actually matters

Four constraints that decide the configuration

The stage is the instrument

Extinction angles and retardation are read off the stage. A 360° rotating stage with 1° graduation, 6′ accuracy and a lock is the minimum for quantitative work; a graphite-spray surface keeps the specimen flat and reduces stray reflection.

Strain-free optics or nothing

Every lens in the path can introduce its own birefringence. Strain-free plan objectives, a strain-free condenser and a centred polarizer/analyzer pair keep the measured retardation attributable to the specimen.

Conoscopy needs a Bertrand lens

Interference figures require a Bertrand lens and a focusable conoscopic path. Both STCX43P and STCX43MP include it, so optic sign and 2V work does not need a second instrument.

Opaque and transparent in one session

Ore mounts, slags and metals are opaque; thin sections are transparent. STCX43MP carries transmitted polarizing and reflected bright-field paths on one stand — the same stage geometry for both, so results are comparable.

Before you request a quote

Four questions we will ask

  • Transmitted, reflected or both
  • Quantitative angle measurement requirement
  • Conoscopy and compensator needs
  • Sample format — thin section, grain mount, polymer film