From a cut and mounted specimen to a traceable grain-size number: the microscope, the stage and the calibration have to agree.
Upright stands are the default for mounted specimens up to about 40 mm tall; inverted stands take heavy or irregular parts directly on the stage, with no height limit beyond travel. If your lab cuts large castings or weld sections, an inverted stand removes a preparation step every time.
Non-metallic inclusions, pits and fine scratches are low-contrast in bright field. Dark field turns them into bright scatterers against a dark background — which is why BD configurations, not BF-only stands, are the standard for inclusion rating to ISO 4967 / ASTM E45.
A photomicrograph is not a result. Grain size (ASTM E112), coating thickness and porosity all need a calibrated scale tied to the objective in use. A coded turret removes the most common source of error in metallographic reporting: the wrong calibration applied to the wrong objective.

Bright-field apochromat optics on an inverted stand — large, heavy or irregular specimens sit on the stage without pre…
View specifications
A dedicated bright/dark-field switching unit brings inclusions, coatings and fine scratches into relief without changi…
View specifications
Bright field, dark field, polarized light and DIC selected from a single illuminator — no hardware swap between contra…
View specifications
CFI60 infinity optics with TU Plan Fluor BD objectives — edge-to-edge flatness, high NA and long free working distance…
View specificationsMaterial, size, feature of interest and the measurement you need to report.